A few thoughts on pressure-mapping inhalers… and Flohms
Currently there’s no agreed standard method or practice to determine the airflow resistance of an inhaler, leaving scientists and engineers to their own devices (sorry!).
Currently there’s no agreed standard method or practice to determine the airflow resistance of an inhaler, leaving scientists and engineers to their own devices (sorry!).
However, this is an important parameter to measure as accurately as possible. It’s essential to determine Qout (the flowrate through the inhaler device corresponding to a “nominal” 4 kPa pressure drop across it) in a repeatable and reproducible manner.
If this measurement isn’t performed correctly, in vitro data may be i) inaccurate and / or ii) more variable than desired; devices that are actually within specification may wrongly be failed, and in the worst case a potentially life-saving medical product may never reach the patient.
So what is a sensible method to use?
To accurately measure the airflow resistance of an inhaler it’s necessary to achieve a number of steady-state flows through it, whilst recording the pressure drops across it that correspond to those flowrates. Ideally, the range of pressure drops should be representative of the likely pressure drops achieved in actual use – e.g. between 0.2 kPa and approximately 10 kPa.
The closest methodology (in terms of experimental setup) provided by the United States Pharmacopeia (USP) is given for the determination of Delivered Dose Uniformity. However, particularly with high resistance inhalers, this method is not ideal, as replacing a high resistance inhaler with a low resistance flowmeter will change the flowrate through the system.
It is far more practical and indeed more accurate to keep the flowmeter in the fluid circuit throughout the duration of the measurements, with a HEPA filter placed upstream to ensure any particulates emitted from the inhaler cannot enter the sensitive flowmeter, where they could deposit on the thermal film sensor and affect measurement accuracy. A simple schematic of the suggested experimental configuration is shown below:

Providing the following testing requirements are satisfied, then the measurement of the airflow resistance will be sufficiently accurate:
Recommended Test Procedure
A few notes on the template
So what’s all this about Flohms?
I’ve worked in the field of respiratory devices for perhaps a little too long, and the units of airflow resistance have always totally annoyed me. They’re so cumbersome!
I’ve talked a few times about renaming the units of resistance, √Pa min L ˉ¹, to “Flohms” (“Flow”-“ohms”) – or even more simply FΩ. This is pure laziness, as I haven’t the time to keep writing it out in full.
I hope you understand…
Looking for industry insights? Click below to get our opinions and thoughts into the world of medical devices and healthcare.